- Model:EA1000VX
- Product:X-ray Fluorescence Analyzer
♦ High-end model for RoHS application
♦ High Count Rate Detector (Vortex®)
♦ Hazardous Substance Measurement Software
Ver.2
High speed measurement
Measurement time has been dramatically improved over standard instruments: one tenth for metal samples and one third for plastic samples compared with the conventional model SEA1000AII. This results in far greater overall throughput.
Cd, Pb, Hg, Br, Cr in plastics | Cd, Pb, Hg, Cr in high concentration BrSb plastics | Cd, Pb, Hg, Br, Cr in brasses | |
EA1000VX | Approximately 30 sec | Within 120 sec | Within 200 sec |
SEA1000AII | Approximately 100 sec | Within 360 sec | More than 2000 sec |
High Count Rate Detector (Vortex®)
The EA1000VX is equipped with the high count rate Vortex detector.
Material ID Function
The EA1000VX rapidly identifies materials of different types of samples as soon as the measurement starts. Previously, an analytical recipe had to be selected first, but now there is no need to select an analytical recipe; the instrument can automatically identify a material.
Improved operability - Operation Panel
The user can monitor the measurement status via the progress bar on the Operation Panel as well as the sound indicators.
New Version of Hazardous Substance Measurement Software with a Variety of New Functions
Software manages the data (retrieval, browsing, analysis, editing, printing, and report preparation) measured by one or more instrument, allowing higher inspection efficiency and cost reduction.
Standard reference materials for various environmental regulations (optional)
A variety of standard reference samples developed and manufactured in-house are available; those include not only for the RoHS restricted elements (Cd, Pb, Hg, Br, and Cr), but also for other elements such as chlorine (Cl), antimony (Sb) and tin (Sn) and so forth.
Item | Specification |
---|---|
Origin | Japan |
Elements | Al13 to U92 |
Sample type | Solid, Powder, Liquid |
X-ray source | X-ray tube (Rh target) Voltage: 15 kV, 40 kV, 50 kV Current: 1 mA (10 μA to 1,000 μA variable) |
X-ray irradiation direction | Bottom-Up method |
Detector | Vortex® (SDD) (Liquid nitrogen not required) |
Analysis Area | 1 mm, 3 mm, 5 mm (automatic switching) |
Sample imaging | Color CCD camera |
Filters | 5 mode automatic switching (including OFF) |
Sample chamber | 370(W)×320(D)×120(H) mm |
Size | 520(W)×600(D)×445(H) mm |
Weight | 60Kg |
Software | X-ray Station
|
Power usage | AC100~240V (50/60Hz) / 210VA |